Syringe pump TOP-5300

Company Name: TOP

Origin: japan

Trade Mark: TOP

Model: TOP-5300

Specifications:

Technical Specification:
1) power SUPPLY:
• AC power source : AC 100 to 240 V ±10% 50/60Hz(approx. )
• ExtrenalDC power source : DC 12.7 to 15V (exact )
• Built-in battery : Rechargeable, DC 12V 700mAh NI-CD Model : BP-53
• Continous operation : Approx. two hours ( when a new battery is fully Recharged with 5mL/h set)
2) power consumption external :
• AC power source : 0.3A ( approx.)
• External DC power source : 0.5A ( exact )
3) Applicable syringes
10mL, 20mL, 30mL and 50mL syrings from
1) TOP, TERUMO, B-D or B. BRAUN
2) B-D, TERUMO, MONOJECT or B.BRAUN can be used ( in the auto selecation [*] mode, only the 50 mL syringe can be used ).
4) flow rang
• 0.1-400.0 mL/h (10mL syringe)
• 0.1-700.0 mL/h (20mL syringe)
• 0.1-900.0 mL/h (30mL syringe)
• 0.1-1500.0mL/h (50mL syringe )
• ( setting Adjustable in 0.1mL increments.)
5) total infusion volume range
• 0.1-999.9 mL
6) infusion accuracy
Brand – specififed mode
• Mechanical accuracy : ±1%
• Accuracy including the syring : ±3%
Auto selection [*] mode(only 50ml syringe can be used.)
• Mechanical accuracy : ±1%
• Accuracy including the syringe :±3%
҉ Accuracy obtained when infusion is performed for over one hour
at an infusion rate of 1.0mL/h or more.
7)Bolus rate
• Approx 400.0mL/h (10mL syringe)
• Approx 700.0mL/h (20mL syringe)
• Approx 900.0mL/h (30mL syringe)
• Approx 1500.0mL/h (50mL syringe)
4- step selection
• Extra high : 119.7±40.0kpa/ 900±300mmHg/ 1.22±0.41 kgf/ cm2
• High : 93.1±33.3kpa/ 700±250mmHg/ 0.95±0.34 kgf/ cm2
• Middle: 66.5±26.7kpa/ 500±200mmHg/ 0.68±0.27Kgf/cm2
• Low : 39.9±20.0Kpa/ 300±150mmHg/ 0.41±0.20Kgf/ cm2
9)Alarm
Low volume (inactive in the auto selection [*] mode), occlusion, lowbattery, syringebarel/ syring plunger clamp disengaged, power source switch, remainder, CPU error.
10) Special functions
• Operation Reminder:If the system is left in the suspend state for approximately three minutes, a buzzer will sound.
• Auto power off: If the system is left in the "Operation Reminder" state for another three minutes while ti is powerd by an internal battery, it will be automatically shut off .
• Alarm Repeat: If two minutes have elapsed without resetting the alarm after it is silenced the alarm sounds again.
• Data memorize : Infusion rate set value/Total infusion volume storage setting is allowed.
• Buzzer setting : Presence/Absence of wait tone and operation tone can be speficied.
• Power source switch alarm : Notifies by the alarm sound that power is no longersupplied from the AC or DC power source and the system has entered the battery operation mode.
11) operation conditions :
• Ambient temperture : 10 to 40 °C
• Relative humidity : 30 to 85% ( no condensation )
• Atmospheric pressure : 70 to106kpa
12) Transportation and storage conditions :
• Ambient temperture : -10 to 45 °C
• Relative humidity : 10 to 90% ( no condensation )
• Atmospheric pressure : 50 to 106kpa
13) classification :
• Class I and Internally powerd equipment, Type CF♥, IPX1 (drip proof)
14) fuse : T1A
15)External dimensiobs / weight :
• Approx 320(W)x116(H) X 139(D)mm / approx 2.2kg

Download Syringe pump TOP-5300