Our Products

infusion pump TOP-3300

Company Name: TOP

Origin: japan

Trade Mark: TOP

Model: TOP-3300

Specifications:

Technical Specification:
1) pump type
• Peristaltic finger
2) Applicable IV sets
• TOP IV sets H- Series ( TOP- H ) for infusion Pumps, and other standard IV sets (15,19,20,and 60 drops/mL)
3) Flow rate
• 1.0-999mL/h(TOP-H series dedicated IV set)
• to 99.9 mL/h, 0.1 mL/h increments; above 100 mL/h, in 1 mL/h increments
• 1.0-400 mL/h (standard 15 drops/mL IV set)
• 1.0-380 mL/h (standard 19 drops/mL IV set)
• 1.0-360 mL/h (standard 20 drops/mL IV set)
• 1.0-150 mL/h (standard 60 drops/mL IV set)
• 4)volume limit range : 1 to 9,999 mL in 1 mL increment
5) infusion accuracy
• ±5% when TOP- H sets are used
• ±10% when other common IV sets are used
6) Alarms
• Empty container, Occlusion Door open Air-in-line, Low battery, Infusion complete, Power source switched, Operator error, and other deviced malfunctions
7)Air bubble detection
• Can be set to any or 3 levels, 50µL 0.1mL/15min, 1mL/15min
8)Occlusion detection pressure
Can be set to any of 4 levels when TOP - H sets are used :
• P1-4: Occlusion not detected
• P1-3: ≤ 160.0 kpa ( ≤ 1,200mmHg/ ≤1.63Kgf/cm²)
• P1-2: ≤ 120.0 kpa ( ≤ 900mmHg/ ≤ 1.22Kgf/cm²)
• P1-1: ≤ 80.0 kpa ( ≤ 600mmHg/ ≤0.82Kgf/cm²)
• P1-0: ≤ 40.0 kpa ( ≤ 300mmHg/ ≤0.41Kgf/cm²)
Can be used to either of 2 levels when other common IV sets are used :
• P2-1: ≤ 160.0 kpa ( ≤ 1,200mmHg/ ≤ 1.63Kgf/cm²)
P1-0: ≤ 80.0 kpa ( ≤ 600mmHg/ ≤ 0.82Kgf/cm²)
9)Special functions
• Automatic Shutoff: When the infusion pump is operating on battery power and has been left for 3 minutes with the infusion stopped or in an alarm state, the buzzer sounds. After the buzzer has been left sounding for 3 minutes, the power turns off automatically to conserve the battery.
• Alarm repeat: Unless the cause of an alarm is corrected within 2 minutes after the operator has pressed the "silence" buzzer key, the buzzer will sound again to remind the opertor of the problem.
• Data memorize function: Retains the most recent flopw rate, volume limit or total volume infused for the next used.
• Sound setting: Audible indication of standby and infusion in progress can be turned on or off.
• Power Source Switch alarm: An alarm sounds when AC power supply has stopped and the unit has switched over to battery operation.
10)Operating conditions
• Ambient temperature: +5 to 40°C
• Humidity: 20 to 90% (no condensation)
• Pressure: 70 to 106 kPa
11)Storage conditions
• Ambient temperature: -10 to 45°C
• Humidity: 10 to 90% (no condensation)
• Pressure: 50 to 106 kPa
12)power source
• AC 100-240V ±10%, 50/60Hz
• Internal battery: Rechargeable DC 12 V, 700 mAh,
Ni-Cd(type BP- 53).
• Battery life is approx. 2 hour when using a new, fully charged battery at a fluid delivery rate of 25 mL/h.
• External DC power supply: DC 12.7 to 15 V (external power cord optional)
13)Input current
• AC power supply: ≤ 0.3 A~ external DC power supply: ≤0.7A ̿
14) Classification
• Class I, type CF, IPX1 (drip- proof)
15) Fuse:T1A
16) Dimensions:90(W) x 216 (H) x 209 (D) mm
17)Weight :pprox. 2.6 kg

Download infusion pump TOP-3300